Table 13: Actuating signal
ID Function
Unit, resolu-
tion
Command
11 Resolution of controller actuating signal in per mill
– negative value
à
device is cooling
– positive value
à
device is heating
[
‰
]
IN_PV_06
13 Controller actuating signal in watts
– negative value
à
device is cooling
– positive value
à
device is heating
[W]
IN_PV_08
Table 14: Cooling
ID Function
Unit
Command
24 Cooling mode: 0 = off / 1 = on / 2 = automatic
[–]
IN_SP_02
Table 15: Safety
ID Function
Unit
Command
35 Timeout communication via interface (1 – 60 seconds; 0 = Off)
[s]
IN_SP_08
73 Status of Safe Mode: 0 = off (inactive) / 1 = on (active)
[–]
IN_MODE_06
162 Heating output restriction
[%]
IN_SP_13
166 Heating output restriction in Safe Mode
[%]
IN_SP_14
168 Pump pressure set point in Safe Mode
[bar]
IN_SP_15
170 Flow rate set point in Safe Mode, only for device with flow controller
[l/min]
IN_SP_16
172 Pressure limitation in Safe Mode, only for device with flow controller
[bar]
IN_SP_17
Table 16: Control parameters
ID Function
Unit
Command
39 Control parameter Xp
[–]
IN_PAR_00
41 Control parameter Tn (181 = Off)
[s]
IN_PAR_01
43 Control parameter Tv
[s]
IN_PAR_02
45 Control parameter Td
[s]
IN_PAR_03
47 Control parameter KpE
[–]
IN_PAR_04
49 Control parameter TnE
[s]
IN_PAR_05
51 Control parameter TvE
[s]
IN_PAR_06
53 Control parameter TdE
[s]
IN_PAR_07
55 Correction limitation
[K]
IN_PAR_09
V6
Integral Process Thermostats and High-Temperature Thermostats
55 / 198