1.
Press the Enter key to open the menu.
2.
Select the menu items
Setup
Basic setup
Ethernet
Services
LAUDA live
Access configuration
.
The following options appear on the display:
3.
Select the desired option and press the Enter key to confirm.
A check mark is set. The entry has been accepted and the rele-
vant option is selected.
The Access configuration menu enables the user to independently
define the data which may be transferred:
[Op. param. readable] and [Service prm. readable].
A distinction is made between operating parameters (such as the set point/
actual temperature (Tset/Tint/Text)) and the service parameters (internal
device parameters). The operating parameters are essentially the parameters
which are displayed to the user on the device display and can be modified
by the user. In contrast, the service parameters are only available to the
LAUDA Service.
Changing the data in the device from the cloud, e.g. as part of remote
maintenance by the LAUDA Service, is activated on the device side via [Op.
param. writable] and [Service prm. writable].
In addition to measures provided in LAUDA.LIVE for authorizing access
and changing device data, e.g. 2-factor authentication, the device-side
access configuration is used by the user for basic restriction/control of the
LAUDA.LIVE services.
6.23
Safe Mode safety fittings
The Safe Mode enables the user to determine an emergency operating mode
for the device and application in advance. In the event of faults in the process
(e.g. failure of the communication with the control station), it is possible to
switch to this sage operating mode quickly and easily.
The device remains in this safe operating mode until Safe Mode is manually
deactivated.
What happens with activated Safe Mode?
n
The parameters configured in Safe Mode are set.
n
A warning
103 Safe Mode
appears in the display (Home window).
n
The parameters affected by Safe Mode cannot be changed.
n
Active programmers are terminated.
n
The “Set point offset" function is switched off.
n
The Operator privileges of the constant temperature equipment are
assigned to the Master operating unit
privileges in the Master” on page 119. The Operator privileges cannot be
relinquished during the active Safe Mode.
Which events lead to the activation of Safe Mode?
n
communication with the control station being interrupted
n
manually by the user
Configuring LAUDA.LIVE access
Fig. 75: Access configuration
Aim of the Safe Mode
V6
Integral Process Thermostats and High-Temperature Thermostats
123 / 198