Integral XT
102
Starting up
YAWE0028 / 24/06/2016
−
With the menu
Setpoint
the set tem-
perature for the test run can be set.
The bath temperature will oscillate less
than about ±15 Kelvin around the set
temperature.
−
Change the display in the adjacent
window and accept with
.
−
With the menu
Identification
the op-
timal control parameters for internal
control or for the internal control and
the external control can be detected
automatically.
To detect the control parameters for
the external application, a temperature
probe must be connected to the ther-
mostat.
−
With the menu
Actual Parameters
the
actual set control parameters can be
stored in parameter set 9.
After the test run the detected control
parameters will be taken over as con-
trol parameters automatically. If the pa-
rameters found do not fulfil your expec-
tations, the before set parameters can
be restored (
þ
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
30,00°C
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
intern
extern
Pump Menu
End
T
set
T
fix
Status
Setpoint
Identification
Actual Parameters
Store in Set 9