Integral XT
108
Starting up
YAWE0028 / 24/06/2016
Command
Over level handling
−
Open the device parameter menu via the
soft key
Menu .
−
With the cursor keys continue to:
à
Settings
à
Over level handling
.
−
The shown window appears.
−
Select the preferred parameter with
or
and confirm with
.
−
See introduction for details.
7.16.4 High-level warning or alarm
%
‰
3 Sec.
−
Acoustic warning signal sounds for 3 seconds when the liquid level
rises so far that the uppermost switching point of the level sensor
has been reached,
%
‰‰
or in case the warning function as described in 7.16.3 was chosen:
−
The acoustic signal with dual-tone sounds.
−
Warning
WXArn 103
(high level) appears when the expan-
sion vessel contains too much liquid.
−
The
WXArn
flashes by turns with the numeral.
%
‰‰
In case the alarm function as described in 7.16.3 was chosen:
−
The acoustic signal with dual-tone sounds.
−
The red LED
above the fault triangle
flashes.
à
Heater switches off on both poles;
à
Pump and refrigerating unit are switched off electronically.
−
Rectify cause of fault (
þ
9.4).
−
If Alarm: Press Enter key. Warnings disappear automatically when
the cause is gone.
−
Also press this key if the unit has been switched off in the fault
state. Warnings disappear automatically when the cause is gone.
Command
High-level warning/alarm
−
The display shows
Warning. To release press Enter key
Security 3 Level too high
or
Alarm AL 6: Level too high is shown in the display and signifies
that unlocking is only possible on the Master control panel.
Level warning
WXArn
103
°C
Level alarm
AL 6
°C
Pump Menu
End
T
set
T
fix
Over level handling
none
Warning
War. + Heater off
Alarm
Summary of Contents for Integral XT 150
Page 2: ......
Page 4: ......
Page 164: ...Integral XT 162 YAWE0028 24 06 2016 Intentionally left blank ...
Page 166: ......
Page 167: ......