Integral XT
10
Brief operating instructions
/ 24/06/2016
2
Brief operating instructions
These brief instructions shall give you the possibility to operate the unit quickly.
For safe operation of the unit, it is absolutely necessary to read carefully all the instruc-
tions and safety notes!
1. Set up the device or complete the configuration as required (
þ
The device should never be tilted or stood upside down!
Note the connection of the hose joints (
þ
2. Pay attention to pressure sensitive loads (e.g. glass apparatus) with a maximum permissi-
ble operating pressure (
þ
3. Only operate the Integral XT when flow through the external load is possible.
4. Open any shut-off valves in the external loads.
5. Compare the details on the rating label with the mains voltage.
Three-phase device: Ensure a clockwise phase sequence.
Only XT 1850 W Order No. LWP 732 and XT 1590 W Order No. LWP 742:
Check the switch position [400 V; 3/PE; 50 Hz or 440-480 V; 3/PE; 60 Hz] for presence of
mains voltage and frequency. An incorrect setting does not result in any damage, but an
error message occurs
Error 367
(
þ
9.4). With the unit switched off, set the in-
correctly set switch to the correct voltage and frequency values. The switch is fitted on the
back of the unit at the top left, behind the cover panel (
þ
2.3).
6. Only connect device to a socket having a safety earth conductor.
7. Switch on the device by the main fuse-switch on the front panel ("ON =
l
").
Bench-top device
Floor-standing device
8. In the display you see either the current out-
flow temperature , e.g.:
or if the device has not yet been filled:
Act. val. outflow temp.
02%32
°C
Fill device
FiLL
°C