Integral XT
YAWE0028 / 24/06/2016
Maintenance
123
9
Maintenance
9.1
Cleaning
9.1.1 Cleaning the surface of the device
Withdraw the equipment mains plug before cleaning.
Cleaning can be carried out with water to which a few drops of surfactant (washing-up liquid) have been
added and using a damp cloth.
No water must enter the control section.
+
Carry out appropriate decontamination if hazardous material is spilt on or in the equip-
ment.
The cleaning or decontamination method is determined by the user's specialist
knowledge respectively the corresponding data sheets. In case of doubt contact the
manufacturer of the hazardous material.
9.1.2 Cleaning the hydraulic circuit
Refer to cleaning procedure (
þ
9.1.3 Draining the water-cooled condenser
+
Important:
With the risk of frost (e.g. transport in winter) drain the condenser on water-
cooled devices.
XT 250 W:
To do this, heat up the outflow to about 20 °C.
Remove the water hose on the water tap. Set the
setpoint to, for example, 10 °C and immediately
after the compressor start-up, blow into the water
return hose with compressed air. Continue until all
water has flowed out of the device. Switch the
device off immediately.
From XT 350 W:
To do this, select the condenser draining mode (
þ
7.7). Remove the water hose on the water tap.
Blow into the water return hose with compressed
air. Continue until all water has flowed out of the
device. Switch the device off immediately.
9.2
Device status
The process thermostat Integral XT can be conveniently checked with the Command remote control.
9.2.1 Interrogating the device type
Menu
à
Settings
à
Device status
à
Device type
With low temperature thermostats the device type is detected automatically and cannot be changed.