Integral XT
YAWE0028 / 11/11/2016
Technical data
153
Table 3 Process thermostats
XT 750 (S)
XT 750 H(S)
XT 950 W(S)
XT 1590 W(S)
XT 1850 W(S)
Operating temp.- ACC range
°C
-50 – 220
-50 – 300
-50 – 220
-90 – 220
-50 – 220
Ambient temp. range
°C
5..40
Humidity
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to 50 % relative
humidity at 40 °C
Device distance to the sur-
roundings
front
cm
50 50 20 20 20
back
cm
50 50 20 20 20
right
cm
50 50 20 20 20
left cm
50
50
6
6
6
Storage temperature range
°C
-20 – 44
the condenser must be completely emptied on a water-cooled device (
9.1.3)
Setting resolution
°C
0.01
Display resolution
°C
Master: 0.01 Command: 0.1 / 0.01 / 0.001
Display accuracy
0.2 °C can be calibrated additively
Filling volume, minimum
L
5.0
5.3
5.0
10.5
9.0
Additional filling volume in the
expansion vessel
L 6.7
6.7
6.7
17.4
17.4
Refrigerant
R-404A
R-404A & R-508A
R-404A
Cooling refrigerating unit
Air
Water
Cooling air temperature
range without performance
loss
°C
10 – 20
10 – 40
Cooling water connections
---
R¾” A
R1” A
minimum diameter of the
cooling water hose
mm ---
---
13
19
19
Cooling water temperature
range /
without performance loss
°C
---
---
10 – 30
10 – 15
10 – 30
10 – 15
10 – 30
10 – 15
Cooling water pressure
bar
---
---
3 – 10
3 – 10
3 – 10
Cooling water consumption:
temperature 15 °C, pressure
3bar
L/h
--- ---
1300
1500
1300
Cooling power
at
20°C ambient t
e
mperatu
re; cooling
water temp
eratu
re 15°C; cooling
water p
ressure 3
bar;
pump level 4 ; if not different n
o
ted
Thermal
transfer
oil
300 °C
KW
---
5.50
---
---
---
200 °C
KW
7.00
7.00
9.00
15.00
18.50
100 °C
KW
7.00
7.00
9.00
15.00
18.50
Ethanol 20
°C KW
6.70
6.70
9.00
15.00
18.50
Ethanol 10
°C KW
6.10
6.10
7.50
13.00
12.50
Ethanol 0
°C KW
4.80
4.80
6.60
10.50
10.30
Ethanol
-10
°C
KW
3.40 3.40 4.60 9.20 7.70
Ethanol
-20
°C
KW
2.20 2.20 3.00 8.50 5.90
Ethanol
-30
°C
KW
1.25 1.25 1.70 8.50 3.80
Ethanol -40
°C KW
0.60
0.60
0.90
7.00 2.20
Ethanol -50
°C KW
0.30
0.30
0.35
5.30 1.20
Ethanol -60
°C KW
---
---
---
3.70
---
Ethanol -70
°C KW
---
---
---
1.80
---
Ethanol -80
°C KW
---
---
---
0.90
---
Ethanol -90
°C KW
---
---
---
0.35
---
Temperature stability at
-10 °C, ethanol with external
load
±K
0.05 0.05 0.10 0.30 0.30
L 5
5
5
10
10
Summary of Contents for Integral XT 150