Integral XT
YAWE0028 / 11/11/2016
Technical data
157
Table 4
High-temperature thermostats
XT 4 H
XT 4 HW
XT 8 H
XT 8 HW
Operating temperature/ACC-range
°C
80 - 320
30 - 320
80 - 320
30 - 320
Ambient temp. range
°C
5 - 40
Humidity
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to 50
% relative humidity at 40 °C
Device distance to the surroundings
front
cm
20 20 20 20
back
cm
20 20 20 20
right
cm
6 6 6 6
left
cm
6 6 6 6
Storage temperature range
°C
-20 – 44
the condenser must be completely emptied on a water-cooled device (
9.1.3)
Setting resolution
°C
0.01
Display resolution
°C
Master: 0.01 Command: 0.1 / 0.01 / 0.001
Display accuracy
0.2 °C calibrated additively
Filling
volume,
minimum L
2.6 2.6 2.6 2.6
Additional filling volume in the expan-
sion vessel
L
5.5 5.5 5.5 5.5
Cooling refrigerating unit
Air
Water
Air
Water
Cooling air temperature range without
performance loss
°C
10 - 40
---
10 - 40
---
Cooling water connections
device (outside)
hose (inside)
inch
mm
1/2“
id 19
---
1/2“
id 19
minimum diameter of the cooling
water hose
mm
--- 13 --- 13
Cooling water temperature range /
without performance loss
°C
---
10 - 30 /
10 - 15
---
10 - 30 /
10 - 15
Cooling water pressure
bar
---
3 - 10
---
3 - 10
Cooling water consumption
temperature 15 °C, pressure 3bar
L/h
--- 600 --- 600
Cooling power
at
20 °C ambi-
ent tempera
ture;
cooling water
temperatu
re 15
°C
;
cooling water pre
ssure 3 bar;
pump level 4;
Ultra 350
300 °C
kW
---
16
---
16
Ultra 350
250 °C
kW
---
16
---
16
Ultra 350
200 °C
kW
---
16
---
16
Kryo 55
200 °C
kW
---
16
---
16
Kryo 55
150 °C
kW
---
15
---
15
Kryo 55
100 °C
kW
---
9
---
9
Kryo 55
50 °C
kW
---
2
---
2
Temperature stability at
-10 °C, ethanol with external load
±K
0.05 0.1 0.05 0.1
Liters
5 5 5 5
Heater power / Power consumption
230 V; 50 Hz
kW
3.5 / 3.7
3.5 / 3.7
---
---
208-220 V; 3/PE~60 Hz
kW
---
---
8.0 / 8.8
8.0 / 8.8
200 V; 3/PE~50/60 Hz
kW
---
---
8.0 / 8.7
8.0 / 8.7
200 V; 50/60 Hz
kW
2.65 / 3.2
2.65 / 3.2
---
---
400 V; 3/PE~50 Hz
kW
---
---
8.0 / 8.8
8.0 / 8.8
208-220 V; 60 Hz
kW
2.85 - 3.2 / 3.3 - 3.5 2.85 - 3.2 / 3.3 - 3.5
---
---
Summary of Contents for Integral XT 150
Page 2: ......
Page 4: ......
Page 164: ...Integral XT 162 YAWE0028 24 06 2016 Intentionally left blank ...
Page 166: ......
Page 167: ......