Integral XT
50
Starting up
YAWE0028 / 24/06/2016
7.5
Level display
The level display renders the current liquid level in the expansion vessel visible.
Master
LE
and 2 x
−
Call level display
LE
.
−
The current level indication is displayed (here
5
).
Command
Display in the various windows of the Command remote
control is possible (
þ
Significance of the level indication
0
Low level alarm (
þ
1
Low level warning (
þ
1 – 14
Stable operation possible.
15
High level (
þ
Estimating the filling amount per step of level indication
The range of level indication steps 1 to 15 corresponds to the additional filling volume in the expansion
vessel (
þ
Example:
Additional filling volume in the expansion vessel for Integral XT 150: 5.5 liters.
Volume per level indication step (average): 6 liters / 14 level indication steps = approx. 0.4 liters.
LE 5
°C
Summary of Contents for Integral XT 150
Page 2: ......
Page 4: ......
Page 164: ...Integral XT 162 YAWE0028 24 06 2016 Intentionally left blank ...
Page 166: ......
Page 167: ......