Integral XT
52
Starting up
YAWE0028 / 24/06/2016
Command
Filling mode
The filling window appears automatically when
the level is too low on switching on the device. It
can however also be started manually:
Menu
à
Pump
à
Start filling mode
.
−
Fill with heat transfer liquid as described
above up to Level 4.
7.6.2
Venting
Command
Filling mode
After the filling process there is normally air/gas
in the system which can originate from the follow-
ing sources:
1. Residual air from the hydraulic circuit.
2. Highly volatile constituents of the heat trans-
fer liquid.
Remove residual air from the circuit:
−
Start venting with the soft key Start . F
indication in Master display starts to
flash.
−
Device comes out of the standby mode.
−
Pump runs up automatically to Power
Level 2, and switches off briefly every 45
seconds for improved venting. The heat-
ing and refrigeration unit are switched off.
Follow the pressure and level indications
until the level falls no further, the pres-
sure indication no longer rises and the
degassing symbol is no longer displayed
for more than at least two minutes. This
takes at least a few minutes and can take
more than an hour under unfavorable
conditions with large-volume loads and
high viscosity.
Pump Menu Screen
T
set
Start
Fill mode
Insert heat carrier liquid
T
out
°C
22,
04
Level
4
T
ext
°C
25,
02
Pump
P
int
2,00bar
Step 3
Pump Menu Screen
T
set
Stop
Fill mode
Softkey Stop: End venting
T
out
°C
24,
82
T
ext
°C
25,
02
Pump
P
int
2,00bar
Step 2
Level
4
Pump Menu Screen
T
set
Start
Fill mode
Softkey Start: Run venting
T
out
°C
24,
82
T
ext
°C
25,
02
Pump
P
int
2,00bar
Step 3
Level
4