Integral XT
58
Starting up
YAWE0028 / 24/06/2016
Command
T
set
or T
fix
−
or the soft key T
set
opens the
setpoint window.
−
123.45
is the setpoint which is still active.
The upper and lower limit temperatures
are displayed (device-specific values).
There are three different possible entry
methods:
1. Change the value with the
or
keys. First you vary the 1/10°C values. If
you hold the key pressed longer, then full
degrees change.
2. Enter the complete number with the nu-
merical duo keys and the key for the
negative sign and decimal point.
3. Using
or
, move the flashing cursor
line to the decimal place which you would
like to change and then change it with
or
.
−
Confirm the value with
or quit the
window with
without having made any
changes.
Two other ways of entering the setpoint:
−
With the soft key T
fix
open the
window shown on the left.
−
The setpoints which you last entered are
shown in the right-hand column. In the il-
lustrated screen the last setpoint was
80.0°C.
−
To accept an earlier setpoint, enter the
right-hand column with
and select the
desired value with
, then accept it with
or cancel with
.
−
In the left-hand column, setpoint temper-
atures which are to be used frequently
can be defined as "fixed settings".
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Enter new setpoint:
123,
45
Min: -40.00°C Max:202.00°C
Fixed settings
Recent setpoints
0.00°C
80.00°C
0.00°C
-35.50°C
0.00°C
20.00°C
0.00°C
38.00°C
0.00°C
-35.70°C
0.00°C
0.00°C
0.00°C
0.00°C
0.00°C
0.00°C
Pump Menu
End
T
set
Edit