7.13 Ramp function
With the ramp function temperature changes over any time period can be conveniently entered. This is
especially advantageous with very low temperature changes (e.g. 0.1 °C/day).
Example: From the current outflow temperature (e.g. 242.4 °C) 200 °C of cooling is to occur over 5
days. Then the temperature change is entered as 200 °C and the time as 5 days.
The ramp function is executed until it is manually terminated or until the temperature
limits T
il
(min) or T
ih
(max) described in Section 7.10.2 are attained.
Command
Ramp function
−
Open the list of device parameters using
the soft key Menu .
−
With the cursor keys continue to:
à
Programmer
à
Ramp function
.
−
Confirm with the key
.
−
Enter a positive or negative temperature
value with Temp. change .
−
With Time enter a figure (without time
unit).
−
With Time unit choose between
Sec-
ond(s)
up to
Day(s)
.
−
Under Status the ramp is started
à
Start
or stopped
à
Stop
.
−
When the ramp function is being execut-
ed,
Ramp active
appears in the window
bar.
−
Without manual switch-off the ramp ter-
minates at the latest at T
il
(min) or T
ih
(max).
Pump Menu
End
T
set
T
fix
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Program 1
Program 2
Program 3
Program 4
Program 5
Ramp function
Prog. Optimization
Pump Menu
End
T
set
T
fix
Status
Temp. change
Time
Time unit
Second(s)
Minute(s)
Hour(s)
Day(s)