103
7.15.7 Limiting the heating and cooling power
7.15.7.1
Actuating signal limit
With the actuating signal limit the maximum heating and / or cooling power can be limited. The setting
occurs in percent of the maximum value.
Command
Controller output limit
−
Open the device parameter menu with
the softkey Menu .
−
With the cursor keys continue to:
change
à
Control
à
Controller outp. limit
.
−
The adjacent window appears.
Select
max. Cool
or
max. Heat
with
the cursor key and confirm with
.
−
Enter the desired percentage.
−
Accept the change with
,
−
or quit the window with
without mak-
ing any changes.
+
In addition, the cooling unit can be completely switched off via the operating menu
à
Inter-
faces
à
Smart Cool
.
7.15.7.2
Dynamic limitation of heating power
With the dynamic heating power limitation the heating power can be limited to protect the heat transfer
liquid from overheating on the heater. See also (
þ
Pump Menu
End
T
set
T
fix
Controller outp. limit
max. Cool
max. Heat
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Controller outp. limit
100,
0
Min: 0,0
Max: 100,0