Integral XT
YAWE0028 / 24/06/2016
Starting up
69
7.10.2 Defining temperature limits
With this function it is possible to define a minimum and maximum outflow temperature in the range of
which the device controls as a maximum. When the temperature limits are attained, the heater or the re-
frigerating machine is switched off and a warning output is given. The thermostat controls down (break-
away) already 2 K before the limits. Thus the entry of a setpoint value can be prevented which would
heat or cool the heat transfer liquid too strongly without a device switch-off occurring (cf. the Section
“Overtemperature protection” (
þ
7.16.1)). If, for example, Kryo 30 is used as the heat transfer liquid,
90 °C is the maximum temperature and -30 °C the minimum temperature.
Command
Temp. Limits
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to:
Temp. Limits
.
−
The minimum and maximum tempera-
tures are displayed.
−
T il (min)
is currently active.
−
Select the limit to be changed with
or
and confirm with
.
−
Enter the desired limit temperature.
−
Accept the change with
,
−
or quit the window with
without mak-
ing changes.
Pump Menu
End
T
set
T
fix
T il
(min)
-50.0 °C
T ih
(max)
302.0 °C
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Lower limit (T il)
-50.
0
Min: -50.0 °C Max: 301.0 °C