Integral XT
74
Starting up
YAWE0028 / 24/06/2016
7.10.8 Entering the offset of the external temperature probe
If a deviation is found during the check using a calibrated reference thermometer probe, e.g. from the
LAUDA DigiCal Series, then the offset (the additive part of the characteristic) of the external measure-
ment chain can be adjusted with the following function. The probe of the calibrated reference thermome-
ter must be placed close to the external temperature probe (external Pt100) so that its thermal contact
to the material is as good as the external Pt100.
Command
Calibration
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to:
à
Settings
à
Calibration
à
External Pt100
à
Calibration
.
−
The adjacent window appears.
−
Confirm selection with
.
−
Continue as described in (
þ
7.10.6) for the
internal temperature probe.
7.10.9 Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
Default
−
Open the device parameter menu via
the soft key Menu .
−
With the cursor keys continue to:
à
Settings
à
Calibration
à
external Pt100
à
Default
.
−
The adjacent window appears.
−
Confirm selection with
.
−
Continue as described in (
þ
7.10.7) for
the internal temperature probe.
Pump Menu
End
T
set
T
fix
internal Pt100
external Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
internal Pt100
external Pt100
Calibration
Default