Integral XT
90
Starting up
YAWE0028 / 24/06/2016
7.14 Timer function / Timer (Command)
Using the timer function, the thermostat can carry out an action at a certain time or after a certain wait-
ing period. Actions are: Switching on the thermostat, entering the standby mode or one of the 5 pro-
grams in the programmer.
Command
Clock
Timer 1
Timer 2
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to:
à
Clock
à
Timer 1
,
−
or to
Timer 2
,
−
with the menu
Status
the selected tim-
er is switched
off
or
on
.
−
The standby key does not stop the timer.
Please exercise caution when the thermostat is in standby mode. The thermostat is
not switched off absolutely safely.
A previously activated timer mode could unintentionally start the thermostat again
from the standby mode!
The menu
Function
is used to define
when
an action is executed:
−
Similar to an electronic mains timer,
Week plan
enables two switching
events to be carried out each day. The
cycle is repeated after 7 days.
−
Time absolute
defines a time and a
date on which a once-only action
(switching event) occurs. The time point
is set with
Set time
and with
Set date
.
−
Time relative
defines a waiting period
after which a once-only action occurs.
With
Set time
up to 99h59min can be
entered. ("Set date" is masked out with
this function selection.)
−
If the
Week plan
is activated, in this
window only
Status
,
Function
and
Week plan
are displayed.
Pump Menu
End
T
set
T
fix
Set time
Set date
Timer 1
Timer 2
Format of date
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Pump Menu
End
T
set
T
fix
Week plan
Time absolute
Time relative
Status
Function
Action
Set Time
Set Date