PTT
Installation and operation
YAME0020 / 30.06.11
16
4.3
Checking the filling level
The optimum amount of tempering liquid is
6.5 ml. With an insufficient filling level for the
tempering liquid it can be topped up even with
the beaker in place. To do this the required
amount of tempering liquid is introduced
through the filling hole using the filling syringe.
It is desirable to have a sufficiently high
filling level of tempering liquid for the best
possible heat transfer.
With longer tempering processes visually
check the filling level.
4.4 Switching
off
●
The PTT can be switched off at
any time
by actuating the
switch ("Off" =
O
).
4.5
Standby
mode
The PTT can be put into a
standby
mode. To do this the control knob is held pressed for at least 3 se-
conds.
In this way all electrical components (fans, magnetic stirrer, power electronics for the thermostating unit)
are deactivated.
The
standby
function is used for periods in which the beaker is removed from the system (e.g. when
changing the sample liquid) or for (measurement) phases which require a vibration-free environment.
The device must be put into the standby mode when the temperature sensor is out of the
sample liquid for longer than two minutes in order to prevent overheating of the system and
triggering of the EMERGENCY-OFF function.
Return to the operating mode occurs by again pressing the control knob for at least three seconds.