11.2
Cooling unit
Tab. 10: Cooling capacity
Unit
MC 250
MC 350
MC 600
MC 1200
Cooling capacity (for 20 °C)
kW
0.25
0.35
0.60
1.20
Cooling capacity (for 10 °C)
kW
0.20
0.28
0.50
1.05
Cooling capacity (for 0 °C)
kW
0.15
0.22
0.36
0.75
Cooling capacity (for -10 °C)
kW
0.09
0.16
0.15
0.40
The cooling capacity is measured for a specified temperature of the
heat transfer liquid. These temperature values are shown in brackets.
The ambient temperature for the measurement is 20 °C; ethanol
was used as heat transfer liquid. The cooling water temperature is
15 °C and the cooling water differential pressure is 3 bar for the
measurement of water-cooled devices.
11.3
Refrigerant and filling quantity
The device contains fluorinated greenhouse gases.
Tab. 11: Devices with 230 V; 50 Hz
Unit
MC 250
MC 350
MC 600
MC 1200
Refrigerant
---
R-134a
R-134a
R-134a
R-134a
Maximum filling weight
kg
0.085
0.095
0.295
0.575
GWP
(100a)
*
---
1430
1430
1430
1430
CO
2
equivalent
t
0.1
0.1
0.4
0.8
Tab. 12: Devices with 220 V; 60 Hz
Unit
MC 250
MC 350
MC 600
MC 1200
Refrigerant
---
R-134a
R-134a
R-134a
R-134a
Maximum filling weight
kg
0.08
0.097
0.285
0.63
GWP
(100a)
*
---
1430
1430
1430
1430
CO
2
equivalent
t
0.1
0.1
0.4
0.9
Tab. 13: Devices with 115 V; 60 Hz
Unit
MC 250
MC 350
MC 600
MC 1200
Refrigerant
---
R-134a
R-134a
R-134a
R-134a
Maximum filling weight
kg
0.08
0.095
0.225
0.585
V05REV08
Microcool
53 / 57