Message
Description
Pump blocked
Pump blocked (pump stopped)
Communication with Command / Base interrupted
Command Touch remote control missing or withdrawn during
operation.
Code
Message
Description
Al 1
T ext Pt100 not available
External actual value, Pt100 is not available.
Al 2
T ext analog
External actual value, analog signal is not available.
Al 3
T ext serial
External actual value of serial interface is not available.
Al 4
Input Analog 1
Analog module: Current input 1, interruption.
Al 5
Input Analog 2
Analog module: Current input 2, interruption.
Al 7
Digital Input
Fault on the digital input/switching contact
Al 12
T ext Ethernet
No actual value message via the Ethernet module
Al 14
Communication interrupted
Communication interrupted on the interface module
8.3 Control system warnings
All warnings start with the prefix 0. Two numbers will follow the
prefix. These numbers can be found in the following table.
Code
English Edition
Description
01
CAN OVERFLOW
Overflow at CAN reception
02
WATCHDOG RESET
Watchdog reset
03
TIL LIMIT
til limit active
04
TIH LIMIT
tih limit active
05
HS OVERHEAT
Heater has overheated
06
FREE 5
07
FREE 6
08
INVALID PARAMETER
Invalid parameter in the memory
09
UNKNOWN NODE
Unknown CAN node module connected
10
OLD SWV R
Control system software version is too old
11
OLD SWV S
Protection system software version is too old
12
OLD SWV B
Operating system software version is too old
13
OLD SWV T
Temperature control system software version is too old
14
OLD SWV A
Analog module software version is too old
V6
PRO bath thermostats and circulation thermostats
121 / 156