The cooling output is measured when the heat transfer liquid rea-
ches a certain temperature. These temperature values are specified
in brackets. The ambient temperature for the measurement is 20 °C
and ethanol was used as a heat transfer liquid. To measure water-
cooled devices, the cooling water temperature is 15 °C and the
cooling water differential pressure is 3 bar.
The cooling water must be a minimum of 5
-
10 K cooler than the
ambient temperature of the device to ensure efficient cooling.
Table 38: Cooling water data
Specification
Value
Maximum cooling water pressure
10 bar
Minimum differential pressure of
cooling water (input - output)
r
p 0 bar
Maximum differential pressure of
cooling water (input - output)
r
p 3.0 bar
Cooling water temperature
15 °C recommended, 10 to 30 °C
permitted (in upper temperature
range with reduced cooling output)
Diameter of cooling water hoses
10 mm
Cooling water hoses
Hoses with an inner diameter of 10 mm are recommended for the
nipples supplied with the device.
Table 39: Single-stage refrigerating machine
Unit
RP 2040
RP 3035
RP 2045
Pump level
Cooling capacity
at 20 °C
W
800
800
1500
8
at 10 °C
W
800
800
1430
8
at 0 °C
W
800
800
1170
8
at -10 °C
W
600
580
840
8
at -20 °C
W
400
350
520
4
at -30 °C
W
190
160
280
4
at -35 °C
W
110
100
200
4
at -40 °C
W
60
---
130
4
at -45 °C
W
---
---
70
4
at -50 °C
W
---
---
---
---
V6
PRO bath thermostats and circulation thermostats
134 / 156