Chemical samples, for example, are placed on special racks, platforms and
rising platforms for temperature control purposes.
NOTICE!
Risk of bath vessel corroding due to different electrochemical
potentials from metal rack and bath
Device damage
If the stainless steel vessel makes direct contact with a rack,
it may result in electrochemical oxidation. Avoid using racks
that may react in this way to direct contact with the vessel.
Use LAUDA stainless steel racks or commercially available
racks made from temperature-resistant plastic.
4.8 External consuming unit
4.8.1 Hoses
CAUTION!
Risk of external hydraulic circuit bursting
Scalding, cold burns
Use hoses with a greater compressive strength than the
maximum possible pump pressure.
CAUTION!
Risk of heat transfer liquid escaping
Scalding, cold burns
The temperature and media resistance of the hoses used for
the device overflow must be suitable for the application.
CAUTION!
Risk of heat transfer liquid escaping due to the use of unsuitable
hoses
Scalding, cold burns
The temperature and media resistance of the hoses must be
suitable for the application.
CAUTION!
Contact with hot or cold hoses
Hot and cold burns
Use insulated hoses for temperatures below 0 °C and
above 70 °C.
Fig. 22: Connected hoses
V6
PRO bath thermostats and circulation thermostats
52 / 156
Summary of Contents for P 10
Page 146: ...V6 PRO bath thermostats and circulation thermostats 146 156 ...