Proline Kryomats
74
Starting up
YAUE0007 / 2/26/2009
Command
−
Control Parameters
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to:
Æ
Control
Æ
Control Parameters
Æ
Control Parameters
.
−
The adjacent window appears.
Parameter endings:
e = Master controller, f = Slave controller.
−
Where applicable change parameters
marked with (auto) to manual input with
Tv manual/auto
.
−
Select the parameters to be changed with
and confirm with
.
−
Then change the value in the following
settings window and confirm with
.
−
Correction limitation
see introduction.
7.11.2.1
Steps for setting the control parameters for external control
1. Activate external control
Ö
2. Setting the slave controller:
2.1. Set parameters to
auto
; Check for thermostat type and change when necessary (RP....)
Ö
−
Choose heat transfer liquid with low viscosity and high thermal capacity.
Ranking: water, ethanol, water-glycol, oil, Fluorinert®.
−
Set pump level as high as possible,
−
make bath circulation strong and fast,
−
choose hose length as short as possible, i.e. 2 x 1m,
−
choose hose cross section as large as possible, i.e. ½ inch,
−
throughput through the external load as large as possible.
2.2. Xpf setting:
−
when oscillating with short period occur (i.e. 30 seconds)
Î
Xpf lower, otherwise higher,
−
in case of bad thermal coupling and large thermal mass
Î
high (i.e. 2...5, or even higher),
−
in case of good thermal coupling and small thermal mass
Î
low (i.e. 0.2 ... 0.7),
−
when rapid temperature response is required simple internal control should be preferred.
Otherwise select small Xpf (0.05 ... 0.1).
3. Setting the master controller (PIDT1-controller):
−
Start with setting Auto and proceed with Manual only when necessary.
Help Menu End T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Correction limitation
Kpe
0,50
Tne
100
Tve (auto) 83
Tde (auto) 8,3
Xpf
4,0
Prop_E(a) 30