Proline Kryomats
YAUE0007 / 2/26/2009
Starting up
87
Command Explanation
segments receive this tolerance setting.
)
−
For ”_“ use also ” ” (blank character).
−
Response from thermostat ”OK“ or in case of error ” ERR_X“ (RS 485 interface
e.g. “A015_OK” or in case of error ”A015_ERR_X”.).
Permitted data formats:
-XXX.XX -XXX.X -XXX.
-XXX
XXX.XX XXX.X
XXX.
XXX
-XX.XX -XX.X -XX.
-XX
XX.XX XX.X
XX.
XX
-X.XX -X.X -X.
-X
X.XX X.X
X.
X
-.XX
-.X
.XX
.X
7.13.6 Read commands (Data requested from the thermostat)
Command Explanation
IN_PV_00
Read bath temperature (outflow temperature).
IN_PV_01
Indication of the controlled temperature (int./ ext. Pt/ ext. Analogue/ ext. Serial).
IN_PV_03
Read external temperature TE (Pt100).
IN_PV_04
Read external temperature TE (Analogue input).
IN_PV_05
Read bath level.
IN_PV_10
Read bath temperature (outflow temperature)
in 0.001 °C.
IN_PV_13
Read external temperature TE (Pt100)
in 0.001 °C.
IN_SP_00
Read temperature setpoint.
IN_SP_01
Read current pump power stage.
IN_SP_02
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
IN_SP_03
Read current overtemperature switch-off point.
IN_SP_04
Read current outflow temperature limit TiH.
IN_SP_05
Read current outflow temperature limit TiL.
IN_PAR_00
Read current value of Xp.
IN_PAR_01
Read current value of Tn (181 = OFF).
IN_PAR_02
Read current value of Tv.
IN_PAR_03
Read current value of Td.
IN_PAR_04
Read current value of KpE.
IN_PAR_05
Read current value of TnE (999 = OFF).
IN_PAR_06
Read current value of TvE.
IN_PAR_07
Read current value of TdE.
IN_PAR_09
Interrogation of the maximum outflow temperature limit.
IN_PAR_10
Read current value of XpF.
IN_PAR_11
Read current value of TnF (181 = OFF).
IN_PAR_12
Read current value of TvF.
IN_PAR_13
Read current value of TdF.
IN_PAR_14
Interrogation of the setpoint offset.
IN_DI_01
Status of contact input 1: 0 = open/ 1 = closed.
IN_DI_02
Status of contact input 2: 0 = open/ 1 = closed.
IN_DI_03
Status of contact input 3: 0 = open/ 1 = closed.
IN_DO_01
State of Contact output 1: