Proline Kryomats
106
Technical data
YAUE0007
/
2/26/2009
11
Technical data and diagrams
The figures have been determined according to DIN 12876.
RP 3050 C
RP 3050 CW
RP 4050 C
RP 4050 CW
Operating temp.- ACC range
°C
-50...200
Ambient temp. range
°C
5...40
Relative humidity
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to 50 % relative
humidity at 40 °C
Temperature range for
storage
°C -20…44
at water-cooled devices the condenser must be completely empty (
Ö
Setting resolution
°C
0.1 / 0.01 (Master); 0.01 (Command)
Display resolution
°C
Master: 0.01 Command: 0,1 / 0,01 / 0,001
Display accuracy
±0.2 °C can be calibrated additively (
Æ
Temperature stability
±K
0.05
Safety equipment
Class
III, FL suitable for flammable and non-flammable liquids
Maximum cooling water
consumption: temperature
15 °C, pressure 3 bar
~
[L/h]
700 700
Water-cooling connections
(DIN EN 10226-1)
G
¾” G
¾”
Heater power
400 V; 50 Hz
kW
maximum 3.5
with heat
transfer oil
200 °C
kW
Data coming soon
Data coming soon
5.00
7.00
with ethanol
20 °C
kW
5.00
6.00
0
°C
kW
3.0
3.5
-20
°C
kW
1.6
1.8
-30
°C
kW
1.0
1.1
-40
°C
kW
0.5
0.6
-50
°C
kW
0.25
0.25
-60
°C
kW
-70
°C
kW
-80
°C
kW
Cooling power
at
20°C t
am
b
@ bat
h temp. (Pump
L
e
vel
6)
-90
°C
kW
Pump type
Pressure pump, 4 power levels (level 5 to 8)
Discharge pressure max.
bar
0.5 at pump power level 8
Flow rate max. (pressure)
L/min
19 at pump power level 8
Hose connections
Thread M16x1; olives 13 mm external diameter
Bath volume from...to
L
23...31
32...44
Bath opening B x L
mm
350 x 200
350 x 350
Bath depth / usable depth
mm
250
Height to top of bath
mm
905
Overall dims. B x L
mm
600 x 700