7.6.2 Defining temperature limits
With this function, it is possible to define a minimum and a maximum temperature in which the
thermostat controls. By reaching the temperature limits, a warning appears. In this way setpoint input
can be prevented which may damage the heat transfer liquid or the apparatus. For example, if water
were used as the heat transfer liquid, +95 °C would be practicable as the maximum temperature and +5
°C as the minimum temperature.
Command
−
Limits
−
Open the device parameter menu via the
soft key Menu .
−
With the cursor keys continue to
Limits
.
−
The minimum and maximum
temperatures are displayed.
−
T il (min)
is currently active.
−
Select the limit to be changed with
or
and confirm with
.
−
Enter the desired limit temperature.
−
Accept the change with
,
−
or quit the window with
without making
changes.