Programmer (PGM only Command)
Almost any temperature/time profile can be created with the programmer. A desired bath temperature
can be approached as quickly as possible or via a defined ramp. Furthermore, the pump level and the
behavior of the switching outputs can be defined. Five temperature/time programs are provided for free
programming. Each program consists of a number of temperature/time segments. Also included are
details of how often the program is to be executed (loops). The sum of all segments of all programs
may be up to a maximum of 150. A warning is given if the creation of more than 150 segments is
attempted.
Typical segments are:
Ramp:
If a time is specified, then the segment is a ramp, which is described by the target temperature,
i.e. the temperature at the end of the segment, and the duration from the start to the end of the
segment.
Step:
Without any specified time the final temperature is approached as quickly as possible.
Temperature hold phase:
No temperature change (i.e. the temperatures at the start and end of a
segment are the same).
)
The programmer can be controlled or changed via the RS 232 interface, the timer or
switching contacts.
7.8.1 Program example
Cooling time depends on
type of device, load, etc.
Real program example with 6 segments
No
T end °C
Time [h:m] Tolerance
No
Pump
Out 1
Out 2
Out 3
Start 30.00°C -------
0.00°C
Start ------- ------- -------
-------
1 30.00°C 00:20
0.10°C 1
2
-------
-------
-------
2 50.00°C 00:20
0.00°C 2
3
-------
-------
-------
3 70.00°C 00:40
0.00°C 3
4
-------
-------
-------
4 70.00°C 00:10
0.10°C 4
2
-------
-------
-------
5 60.00°C 00:30
0.00°C 5
2
-------
-------
-------
6 30.00°C 00:00
0.00°C 6
2
-------
-------
-------