ECOLINE Staredition

 

 

 

27.06.2018

/

YACE0068

 

Description

 

Peculiarities of the types RE 212 J, RE 312 J: 

Description 

The thermostats are especially equipped to calibrate glass thermometers or electric thermometers. The 
cylindrical (working) chamber can be adjusted in its height up to 20 mm approximately and therefore it is 
possible to adjust the surface of the liquid bath within the working chamber higher than the (bath) cover. 

Therefore, the total immersing thermometers can be read right at the point of immersion. Additionally the 
separate working chamber provides a constant immersion depth, independent from the volume 
extension of the bath liquid, an extremely good temperature accuracy and distribution of temperature. 
There are pump connectors to connect other external closed circuits, but won’t be available if the 
thermostat is used for calibrating. 

 

 
 
 
 
 
 
 

Side view 

Top view 

 

EU conformity

  The device complies with the basic health and safety requirements outline in the 

Directives listed below.  

  Machinery Directive 2006/42/EC  

  EMC Directive 2014/30/EU  

LAUDA DR. R. WOBSER GMBH & CO. KG 

– Pfarrstraße 41/43 –  

97922 Lauda-Königshofen 

– Germany 

 

 

 

The device does not fall under Pressure Equipment Directive 
2014/68/EU because the device is only classified as high as Category 1 
and is covered by the Machinery Directive. 

Connection tube 

LZM 045 

Screw for adjustment (3x) 

Summary of Contents for RE 212 J

Page 1: ...Lauda Königshofen Germany release 06 2018 c replaces release 11 2016 V02 08 20156 b1 07 04 YACE0068 Phone 49 0 9343 503 0 Fax 49 0 9343 503 222 e mail info lauda de Internet http www lauda de Operating instructions Calibration thermostat RE 212 J RE 312 J Annex to manual RE 212 RE 312 YACE 0060 YACE 0065 ...

Page 2: ......

Page 3: ... immersion depth independent from the volume extension of the bath liquid an extremely good temperature accuracy and distribution of temperature There are pump connectors to connect other external closed circuits but won t be available if the thermostat is used for calibrating Side view Top view EU conformity The device complies with the basic health and safety requirements outline in the Directiv...

Page 4: ... and fix it that way it rises 3 mm above the bath opening Refill the bath liquid only up to 15 mm underneath the bath opening and only at working temperature and switched off pump Recommendation for the adjustment of the cylindrical working chamber Switch the button for the distribution of the pump flow to ext see diagram Pos 1 Position maximum flow rate in the external circuit working chamber The...

Page 5: ... V 60 Hz kW kW kW 2 25 2 25 1 3 2 25 2 25 1 3 Eff Cooling capacity with Ethanol at ambient temperature 20 C 20 C 0 C 20 C 30 C kW kW kW kW 0 30 0 23 0 13 0 04 0 30 0 23 0 13 0 04 Pump type Pressure pump with selection of 5 power stages Pump output max bar 0 4 Discharge pressure max L min 17 Pump connections for tubing mm Nipples dia 13 mm M16 x 1 Bath filling capacity max L 9 12 Bath opening B x T...

Page 6: ...t This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Instructions for Class A ...

Reviews: