Technical Data

 

Technical Data 

The figures have been determined according to DIN 12876. 

Calibration thermostat 

 

RE 212 J 

RE 312 J 

Working temperature range  

°C 

- 30 

– 200 

Ambient temperature range 

°C 

– 40  

Setpoint resolution 

°C 

0.1 / 0.01 

Display resolution 

°C 

0.05 

0.1 / 0.05 / 0.01 

Display  

 

±0.2 K with additive calibration  

Temperature stability  

±0.01 

Safety feature 

 

 

FL 

Miscellaneous functions 

 

2-line LCD-display 

2-line LCD-display, program 

compilation, external control, 

analog inputs/outputs 

 

 

Automatic compressor, Cooling capacity adjustment 

Heating capacity 

 

 

 

230 V; 50 Hz 

230 V; 60 Hz 

115 V; 60 Hz 

kW  

kW  

kW 

2.25  

2.25  

1.3 

2.25  

2.25  

1.3 

Eff. Cooling capacity @  
with Ethanol  
at ambient temperature 
20 °C 

20 °C  

0 °C  

-20 °C  

-30 °C 

kW  

kW  

kW  

kW 

0.30  

0.23  

0.13  

0.04 

0,30  

0,23  

0,13  

0,04 

Pump type 

 

Pressure pump with selection of 5 power stages 

Pump output max.

 

bar 

0.4 

 

Discharge pressure max. 

 

L/min 

17 

 

Pump connections (for tubing) 

mm 

Nipples dia 13 mm (M16 x 1) 

Bath filling capacity max. 

– 12 

Bath opening (B x T) 

mm 

Ø 150 

Bath depth 

mm 

200 

Working depth 

mm 

180 

Height upper surface of bath 

mm 

441 

Overall dimensions (B x T x H) 

mm 

250 x 400 x 602 

Weight 

kg 

30 

Power consumption 

 

 

 

230 V; 50 Hz  

kW  

2.3 

2.3  

230 V; 60 Hz 

kW  

2.3 

2.3  

115 V; 60 Hz 

kW 

1.4 

1.4 

 

 FL 

compatible for flammable and non-flammable liquids.

 

 

at pumping capacity stage 

 

 

not valid when used for calibrating, because the pressure connection is used for a connecting tube

 

Summary of Contents for RE 212 J

Page 1: ...Lauda Königshofen Germany release 06 2018 c replaces release 11 2016 V02 08 20156 b1 07 04 YACE0068 Phone 49 0 9343 503 0 Fax 49 0 9343 503 222 e mail info lauda de Internet http www lauda de Operating instructions Calibration thermostat RE 212 J RE 312 J Annex to manual RE 212 RE 312 YACE 0060 YACE 0065 ...

Page 2: ......

Page 3: ... immersion depth independent from the volume extension of the bath liquid an extremely good temperature accuracy and distribution of temperature There are pump connectors to connect other external closed circuits but won t be available if the thermostat is used for calibrating Side view Top view EU conformity The device complies with the basic health and safety requirements outline in the Directiv...

Page 4: ... and fix it that way it rises 3 mm above the bath opening Refill the bath liquid only up to 15 mm underneath the bath opening and only at working temperature and switched off pump Recommendation for the adjustment of the cylindrical working chamber Switch the button for the distribution of the pump flow to ext see diagram Pos 1 Position maximum flow rate in the external circuit working chamber The...

Page 5: ... V 60 Hz kW kW kW 2 25 2 25 1 3 2 25 2 25 1 3 Eff Cooling capacity with Ethanol at ambient temperature 20 C 20 C 0 C 20 C 30 C kW kW kW kW 0 30 0 23 0 13 0 04 0 30 0 23 0 13 0 04 Pump type Pressure pump with selection of 5 power stages Pump output max bar 0 4 Discharge pressure max L min 17 Pump connections for tubing mm Nipples dia 13 mm M16 x 1 Bath filling capacity max L 9 12 Bath opening B x T...

Page 6: ...t This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Instructions for Class A ...

Reviews: