Command
Explanation
IN_SP_00
Read temperature setpoint.
IN_SP_01
Read current pump power level.
IN_SP_02
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
IN_SP_03
Read current overtemperature switch-off point.
IN_SP_04
Read current outflow temperature limit TiH.
IN_SP_05
Read current outflow temperature limit TiL.
IN_PAR_00
Read control parameter Xp.
IN_PAR_01
Read control parameter Tn (181 = OFF).
IN_PAR_02
Read control paramete Tv.
IN_PAR_03
Read control parameter Td.
IN_PAR_04
Read control parameter KpE.
IN_PAR_05
Read control parameter TnE (980 = OFF).
IN_PAR_06
Read control parameter TvE (0 = OFF).
IN_PAR_07
Read control parameter TdE.
IN_PAR_09
Read value of correction limitation
IN_PAR_10
Read control parameter XpF.
IN_PAR_14
Interrogation of the setpoint offset.
IN_PAR_15
Read control parameter PropE.
IN_DI_01
State of contact input 1: 0 = open/ 1 = closed.
IN_DI_02
State of contact input 2: 0 = open/ 1 = closed.
IN_DI_03
State of contact input 3: 0 = open/ 1 = closed.
IN_DO_01
State of Contact output 1:
0 = make-contact open/ 1 = make-contact closed.
IN_DO_02
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed.
IN_DO_03
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed.
IN_MODE_00
Keyboard Master: 0 = free / 1 = locked. .
IN_MODE_01
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analogue / 3 = ext. Serial.
IN_MODE_02
Standby operation: 0 = Device ON / 1 = Device OFF.
IN_MODE_03
Keyboard Command remote control: 0 = free / 1 = locked.
IN_MODE_04
Setpoint offset source: 0=normal/1=ext.Pt/2=ext.analogue/3=ext.serial.
TYPE
Read equipment type (response e.g. RP 3050).
VERSION_R
Read software type of control system.
VERSION_S
Read software type of protection system.
VERSION_B
Read software type of Command.
VERSION_T
Read software type of cooling system.
VERSION_A
Read software type of analogue module.
VERSION_V
Read software type of RS-232/485 module.
VERSION_D
Read software type of digital module.
VERSION_M_0
Read software type of solenoid valve (cooling water).
VERSION_M_1
Read software type of solenoid valve (automatic refill).
VERSION_M_3
Read software type of solenoid valve (shut-off valve 1)
VERSION_M_4
Read software type of solenoid valve (shut off valve 2).
VERSION_M_5
Read software type of high temperature cooler.
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......