Proline Kryomats
106
YAUE0007 / 10/2/2018
Command
Explanation
STATUS
Read equipment status 0 = OK, -1 = error.
STAT
Read error diagnosis response:
XXXXXXX
X = 0 no error, X = 1 error.
1. char = error.
2. char = Alarm.
3. char = Warning.
4. char = over temperature.
5. char = low-level error.
6. char = high-level error (at adjustment alarm).
7. char = no external control variable.
RMP_IN_00_XXX
Read a program segment XXX
(response: e.g. 030.00_010.00_005.00_001.00
set point temperature
30.00 °C, time = 10 min, tolerance = 5.00 °C, pump level = 1).
RMP_IN_01
Read the current segment number.
RMP_IN_02
Read the set number of program runs.
RMP_IN_03
Read the current program run.
RMP_IN_04
Read the program to which further instructions apply.
RMP_IN_05
Read which program is currently running (0 = none).
LOG_IN_00_XXXX
Query a measuring point XXXX from data logger
(Reply: e.g. 020.00_021.23_030.50 => set point temperature = 20.00 °C, bath
temperature = 21.23 °C, external temperature = 30.5 °C).
LOG_IN_01
Read all measuring points from data logger
As a difference to the command “LOG_IN_00”, a tabulator is used here as
sepa
rator instead of ,_’ . The measuring points are separated by CR and LF. The
end is marked by CR LF CR LF.
LOG_IN_02
Read the start time from data logger
(Reply: e.g. 20_14_12_20
day 20, time 14:12:20).
LOG_IN_03
Read the acquisition interval from the data logger (Reply in seconds).
For ”_“ use also ” ” (blank character).
The equipment response is always in the fixed decimal format ”XXX.XX“ or for
negative values “-XXX.XX” or ” ERR_X“. (RS-485 interface e.g.. ”A015_
XXX.XX” or ”A015_-XXX.XX” or ”A015_ERR_X”).
The command from the computer must be terminated with CR, CRLF, or LFCR.
The response of the thermostat is always terminated with CRLF.
After each command sent to the thermostat, it is necessary to wait for the reply
before sending another command. This ensures that the sequencing of inquiries
and answers is clear.
CR = Carriage Return (Hex: 0D)
LF = Line Feed (Hex: 0A)
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......