YAUE0007/ 10/2/2018
Safety information
11
Thermal medium hoses and other hot parts must not come into contact with the mains cable.
With the use of thermostats as circulating thermostats hot or cold liquid can be emitted when the hose
breaks, presenting a hazard to persons and material.
If no external load is connected, the pump outflow must be closed (use screw plugs) and the bypass
valve must be set to "internal" (
4.3).
Take into account the thermal expansion of the heat transfer oils with increasing bath temperature.
Irritant vapors may develop, depending on the heat transfer liquid and operating mode used. Always
ensure that the vapors are adequately extracted. Use the bath cover.
By changing the heat transfer liquid from aqueous heat transfer liquid to a thermal transfer liquid for
temperatures above 100°C, carefully remove all water residues, including from the hoses and loads.
When doing this, also open the blanking caps of the pump outputs and inputs and blow
compressed air through all the pump outputs and inputs.
Danger of scald due to delay in
boiling!
Withdraw the mains plug before cleaning, maintenance or moving the thermostat.
Specialist personnel must only carry out repairs in the control section.
The following action may start the thermostat unintentionally from the standby mode: Previously
activated timer mode (
7.10), "Start" command via interfaces (
8).
Figures of temperature constancy and display accuracy apply under normal conditions according to
DIN 12876. Electromagnetic high frequency fields may in special cases lead to unfavorable values.
Safety is not impaired.
Only water-cooled devices:
The return hose of the water-cooling must be securely fixed on the outlet port in order to prevent the
hose sliding off uncontrollably, also during pressure surges.
The return hose of the water-cooling must be fixed on the outlet port that hot cooling water cannot
splash out.
It is essential to prevent kinking or squashing of the return hose for the water cooling. Excessive
pressure can cause the cooling water hoses to tear and hot water to escape.
To prevent damages by a leakage of the cooling water system its recommended to use a leak-water
detector with shut-off valve (Aqua Stop).
1.4
EU conformity
The device complies with the basic health and safety requirements outline in
the Directives listed below.
Machinery Directive 2006/42/EC
EMC Directive 2014/30/EU
LAUDA DR. R. WOBSER GMBH & CO. KG
– Pfarrstraße 41/43 –
97922 Lauda-Königshofen
– Germany
The device does not fall under Pressure Equipment Directive
2014/68/EU because the device is only classified as high as
Category 1 and is covered by the Machinery Directive.
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......