Proline Kryomats
112
Maintenance
/ 10/2/2018
9.1.5
Fault memory
For the analysis and localization of faults the Command version includes a fault memory in which up to
45 fault and alarm messages are saved.
Command
Error store
Settings
Device status
Error
store
Display
.
The last message is at the top.
Each message line can be marked with
the cursor keys. The message appears in
plain text in the footer.
Under “Source”, the CAN node is
displayed which signaled the fault.
Code is the number, which in the Master
is shown in the display until the cause
has been rectified.
Type: Alarm, Warning or Fault (Error).
9.2
Cleaning
9.2.1
Cleaning the surface of the device
Cleaning can be carried out with water to which a few drops of surfactant (washing-up liquid) have been
added and using a damp cloth.
Carry out appropriate decontamination if hazardous material is spilt on or in the
equipment.
The cleaning or decontamination method is determined by the user's specialist
knowledge. In case of doubt, contact the manufacturer.
No. Source Code Type Date
Time
10 Safety 2 Alarm ------------ -----------
9 Safety 4 Warn. 28.08.03 15:32:02
8 Contro. 32 Error 17.07.0310.:52:02
7 Contro. 3 Warn. 06.06.03 11:15:11
6 Contro. 9 Alarm 05.06.03 08:45:01
5 Contro. 3 Alarm 01.06.03 17:58:22
4 Contro. 4 Warn. 28.05.03 20:01:22
3 Contro. 5 Warn. 27.05.03 07:58:00
Low level
Pump Menu
End
T
set
T
fix
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......