114
Maintenance
/ 10/2/2018
9.3.1
Servicing
LAUDA Thermostats largely require no service. If the heat transfer liquid becomes contaminated, it
should be replaced (
6.2).
At the back of the Master head a main fuse switch
is located which
interrupts the mains connection when an overload occurs. It is then in the "
O
"
position and can be set in the "
–
" position again.
If the fuse trips again, Service must locate the cause.
Additionally, a safety fuse, which protects the low voltages, is situated on the
mains board. If a fuse fails (
mains lamp does not light) only replace with a
fuse with the specified data (one x T (= slow-blow) 10 A, size 5 x 20
the Fuse
is located in the unit as shown below).
UL 533
Summary of Contents for RP 3050 C