Maintenance
117
9.3.4.2.2
Decalcifying the water cooling circuit
At regular intervals of 3 months or longer, the water-cooled condenser must be decalcified or cleaned.
This depends on the hardness of the cooling water and the degree of soiling. Drain according to (
Required equipment:
Two containers of approx. 10 to 20 liters volume.
Use a suitable pump (drum pump) or a hose with funnel. Place the funnel as high as
possible so that the device can fill quickly.
Fit connecting hoses between container, pump, cooling water inlet and between cooling
water outlet and back to container.
Change containers when
top one is empty.
Acting time:
Continue the pump stage until most of the foamy reaction, usually at the
start, has decayed. Generally, this is achieved after about 15 to 30 minutes.
Decalcifier:
Water with LAUDA Decalcifier LZB 126.
It is essential to follow the safety instructions when handling the chemicals.
Flushing:
Allow at least 30 liters of water to flow through.
9.3.5
Repair information
If you need to send in a unit for repair, it is essential to first contact the LAUDA Service Constant
Temperature Equipment (
9.5).
When sending in the unit, ensure that it is carefully and properly packed. LAUDA
cannot be held liable for any damage caused by improper packing.
Standard & Alternative
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......