Preparation
YAUE0007 / 10/2/2018
6.3
Connection of the cooling water
Note that the following conditions apply for the connection of the cooling water supply:
Cooling water pressure (feed - outlet)
maximum 10 bar overpressure
Differential pressure (feed - outlet)
minimum 3.0 bar
Cooling water temperature
10 to 15 °C recommended,
10 to 30 °C admissible (with power restrictions)
Cooling water quantity
see Technical Data (
Cooling water hose for connection to the device
minimum 13 mm