Proline Kryomats
42
Starting up
YAUE0007 / 10/2/2018
7.1
Mains connection
Compare the rating on the nameplate (back of control head and behind the front panel) with the mains
voltage.
Connect unit only to a socket with a protective earth conductor (PE).
No liability is accepted for incorrect mains connections.
Ensure that pump connectors without external loads are closed off.
Ensure that the unit is filled according to Section 6.2.
Note for electric installation on site:
Single-phase devices:
Single-phase devices must be protected with a 16 ampere circuit breaker fitted during installation.
Exception: Devices with 13 ampere UK plugs.
Three-phase devices:
For three-phase devices the rating of the circuit breaker must match the power consumption of the
device. The value will be found on the type plate. In each case, select the next higher value. Using an
excessively high rated circuit breaker is not permissible.
7.2
Switching on
Check whether the switch at the back of the master control element
is in the "On =
–
" position.
Check whether the switch at the front of the master control element
is in the "On =
l
" position.
Set the rotary switch on the front panel to "ON =
l
". The unit starts
operating.
The green LED in the master control element for "Mains ON" is lit.
1 s
An acoustic signal is emitted for about 1 second.
It is quite normal if the refrigerating machine makes a rattling sound
for a few seconds.
The unit starts its self-test. All display segments and symbols
appear for about 1 second.
Self-test
88(((
°C
EXT
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......