Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
53
Select desired position with the cursor
keys
(black background).
With the soft key
Edit open the
window shown on the left.
Enter fixed temperature setpoint as
described above and accept into the list
with
or cancel with
.
Select and accept values from the list of
fixed settings as described above for the
“Recent setpoints”.
7.5.2
Displaying the actual external temperature
With all Proline Thermostats an external temperature probe can be connected, which for example....
1. ...can be used as an independent temperature measurement channel.
2. ...can be used as the controlled variable for the bath temperature in applications with a
noticeable temperature gradient (between the internal bath temperature and an external
load). The setup is described in Section 7.5.4. With the function described in the following,
you only change over the display.
External actual temperatures can also be read in by interface modules (
8).
Connection of the external Pt100 to the Lemo socket 10S
Contact on
socket 10S
1
2
3
4
+
+
-
-
I
U
U
I
Current circuit
Voltage circuit
Voltage circuit
Current circuit
Pt100
DIN EN 60751
Plug: 4-pole Lemosa for Pt100 connection (Order No. EQS 022).
Use screened connecting leads. Connect screen to plug case.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Enter new setpoint:
123.
45
Min: -40.00°C Max:202.00°C
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......