Proline Kryomats
54
Starting up
YAUE0007 / 10/2/2018
Master
EXT
or
Switches to the actual-value display of the external temperature
probe (or to the actual value received from an interface module
EXT
is lit in green next to the row of figures.
If no external Pt100 probe is connected,
-----
is displayed.
Command
T
ext
Provided an external temperature probe is
connected, its value is displayed in the
lower left part of the standard and super
windows (applies to the works setting for
the window partitioning).
External actual temperatures can also
be read in via interface modules
(
8).
7.5.3
Setting pump power or standby
With the Proline Varioflex pump, four pump levels (level five till level eight) are available, with which the
bath circulation, flow rate and pressure, the noise generated and the mechanical heat input can be
optimized. Pump level eight gives the best bath circulation and temperature homogeneity.
Master
Pu
and 1 x
Call pump power levels display
Pu
.
The current pump level is displayed (here
5
).
The pump levels display flashes.
or
Select pump level (pump speed = pump power):
5
to
8
for pump operation. Pump responds immediately!
0
activates the standby function (pump, heater and refrigerating
machine are deactivated).
02%02
°C
EXT
Pu 5
°C
Pump Menu Screen
T
set
T
fix
T
int
°C
25.
01
T
set
°C
25.
00
Level
4
T
ext
°C
25.
02
Pump
3
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......