Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
57
Command
Control Variable
Open the device parameter menu with the
soft key Menu .
With the cursor keys, change further to
Control
Control Variable
.
intern
is currently active.
Select other control variables (only
displayed when present) with
or
and confirm with
or End ,
or quit the window with without making
any changes.
or
Change temperature probe source:
1
for the internal probe,
EP
only when an external probe is connected,
EA
only when an analog module is connected and configured,
ES
only when a serial module is connected and is continuously
receiving actual values from a PC.
Wait 4 seconds
or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with Enter key.
If
EP
EA
or
ES
has been selected,
then the green
LED indicates that the control has regulated to
the external temperature signal.
Pump Menu
End
T
set
T
fix
intern
External Pt100
Analog module
Digital module
Control Variable
12!45
°C