Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
65
7.6.5
Setting the volume of the acoustic signals
The LAUDA Proline Thermostats signal alarms as a dual-tone acoustic signal and warnings as a
continuous tone.
Command
Sounds
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Settings
Basic Settings
Sounds
.
Select either
Alarm
or
Warning
.
Example on left:
Alarm
is set to loud.
Select the desired volume with
or
.
Accept selection with
or End or
quit the window with
without making
changes.
7.6.6
Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer (e.g. from the LAUDA DigiCal Series) a
deviation is found, then the offset (i.e. the additive part of the characteristic) of the internal measuring
chain can be adjusted with the following function. The reference thermometer must be dipped into the
bath according to the details on the calibration certificate.
Command
Calibration
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Settings
Calibration
intern Pt100
Calibration
.
The window shown on the left appears.
Confirm selection with
.
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
loud
medium
low
off
Alarm
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......