Proline Kryomats
66
Starting up
YAUE0007 / 10/2/2018
The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with
or End ,
or quit the window with
without making
changes.
7.6.7
Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
Default
Open the device parameter menu via
the soft key Menu .
With the cursor keys continue to
Settings
Calibration
intern Pt100
Default
.
The window shown adjacent appears.
Confirm selection with
.
Confirm the control dialog on the right
with
or cancel with
.
Return to the measurement window with
End or
.
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temperature value of the ref.
temp. measurement device:
20.
15
Min: -50.0°C Max: 302.00
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......