Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
67
7.6.8
Entering the offset of the external temperature probe
If a deviation is found during the check using a calibrated reference thermometer, e.g. from the LAUDA
DigiCal Series, then the offset (the additive part of the characteristic) of the external measurement chain
can be adjusted with the following function. The reference thermometer must be dipped nearly by the
external temperature probe into the consumer bath according to the details on the calibration certificate.
Command
Calibration
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Settings
Calibration
extern Pt100
Calibration
.
The adjacent window appears.
Confirm selection with
.
Continue as described in (
7.6.6) for the
internal temperature probe.
7.6.9
Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
Default
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Settings
Calibration
extern Pt100
Default
.
The adjacent window appears.
Confirm selection with
.
Continue as described in (
7.6.7) for the
internal temperature probe.
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......