Proline Kryomats
70
Starting up
YAUE0007 / 10/2/2018
7.8
Programmer
Almost any temperature/time profile can be created with the programmer. A desired bath temperature
can be approached as quickly as possible or via a defined ramp. Furthermore, the pump level and the
behavior of the switching outputs can be defined. Five temperature/time programs are provided for free
programming. Each program consists of a number of temperature/time segments. Also included are
details of how often the program is to be executed (loops). The sum of all segments of all programs
may be up to a maximum of 150. A warning is given if the creation of more than 150 segments is
attempted.
Typical segments are:
Ramp:
If a time is specified, then the segment is a ramp, which is described by the target temperature,
i.e. the temperature at the end of the segment, and the duration from the start to the end of the
segment.
Step:
Without any specified time the final temperature is approached as quickly as possible.
Temperature hold phase:
No temperature change (i.e. the temperatures at the start and end of a
segment are the same).
The programmer can be controlled or changed via the RS-232 interface, the timer or
switching contacts.
7.8.1
Program example
Cooling time depends on
type of device, load, etc.
Real program example with 6 segments
No
T end °C
Time [h:m] Tolerance
No
Pump
Out 1
Out 2
Out 3
Start
30.00°C
-------
0.00°C Start
-------
-------
-------
-------
1
30.00°C
00:20
0.10°C
1
2
-------
-------
-------
2
50.00°C
00:20
0.00°C
2
3
-------
-------
-------
3
70.00°C
00:40
0.00°C
3
4
-------
-------
-------
4
70.00°C
00:10
0.10°C
4
2
-------
-------
-------
5
60.00°C
00:30
0.00°C
5
2
-------
-------
-------
6
30.00°C
00:00
0.00°C
6
2
-------
-------
-------
Pump
Menu
End
Insert
Delete Pump
Menu
End
Insert
Delete
°C
80
70
60
50
40
30
20
10
0
20
40
60
80
100
120
t/min
Start
This is Segment No. 2
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......