Each program begins with the segment "Start". It defines at which temperature
Segment 1 is to continue the program. It is not possible to specify a time for the Start
segment. For thermostats without cooling ability, the start temperature must be
selected higher than the bath temperature, which prevails before the program start.
Without the Start segment, Segment 1 would be different depending on the bath
temperature at the start of the program.
Edited program example (see dashed curve in the graph on previous page).
No
T end °C
Time [h:m]
Tolerance
No
Pump
Out 1
Out 2
Out 3
Start
30.00°C
-------
0.00°C
Start
-------
-------
-------
-------
1
30.00°C
00:20
0.10°C
1
2
-------
-------
-------
2
50.00°C
00:20
0.00°C
2
2
-------
-------
-------
3
50.00
°C
00:20
0.10°C
3
2
-------
-------
-------
4
70.00°C
00:20
0.00°C
4
2
-------
-------
-------
5
70.00°C
00:10
0.80°C
5
2
-------
-------
-------
6
60.00°C
00:30
0.00°C
6
2
-------
-------
-------
7
30.00°C
00:00
0.00°C
7
2
-------
-------
-------
Pump
Menu
End
Insert
Delete
Pump
Menu
End
Insert
Delete
Insert new segment (
Section 0)
Change segment time or tolerance (
Section 0)
The field tolerance
(refer to the above program table and the graph below):
It facilitates exact conformance to the dwell time at a specified temperature. Segment
1 is not processed until the bath temperature is within the tolerance range
, so that
the ramp (Segment 2) starts delayed at
.
A tolerance range which is too tight can however also cause undesired delays.
In
particular with external control
the range should not be chosen too tightly. In
Segment 5 a larger tolerance has been entered, so that the desired time of ten
minutes is maintained even with settling action
.
Only flat (slow) ramps should be programmed where necessary with a tolerance
range. Steep ramps which lie close to the maximum possible heating or cooling rates
of the thermostat may be severely delayed by a tolerance range that is too tight (here
in Segment 2)
.
Example for the influence of the tolerance field input
in case of external bath temperature control:
The setpoint temperature of the programmer is
shown in grey.
The actual temperature in the external bath
container is represented as a continuous line.
°C
80
70
60
50
40
30
20
10
0
20
40
60
80
100
120
t/min
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......