Proline Kryomats
76
Starting up
YAUE0007 / 10/2/2018
A new segment is produced by moving
the cell with the black background to a
blank line with the cursor keys and then
pressing the soft key
Insert . The
values of the cell located above it are
automatically copied.
If the field in the column
T end °C
has a
black background, the entry mode "End
of segment temperature" is obtained by
pressing the
key. Depending on the
setting, that is the temperature, which the
thermostat is to achieve on the internal or
external temperature probe.
Enter the value, confirm with the
key
and continue to the "Time" entry field with
.
If the field in the column
Time
" has a
black background, the entry mode for the
"Segment time" time setting is obtained
by pressing the
key.
If 0 is entered into the field "Time",
“------”
appears. Then the final temperature is
approached as quickly as possible. With
a time entry the final temperature is
obtained exactly after the time expires
(ramp).
Enter the segment time and confirm with
the
key.
Continue to the "Tolerance" entry field
with
.
If the field in the column "Tolerance" has
a black background, the entry mode for
the "Temperature tolerance" is obtained
by pressing the
key. It defines how
accurately the end of segment
temperature is to be obtained before the
next segment is processed.
A tolerance which is selected too small
can stop the next segment from being
started according to plan.
Set the temperature tolerance and
confirm with
.
Continue with
to the entry field
"Pump".
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1 2
3
4
5
Temp. tolerance (0=off):
10.
00
Min: 0.00°C Max:450.00°C
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
End of segment temperature:
25.
00
Min: -150.00°C Max:450.00°C
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Input segment time:
003:00
Hours(max.999):Minutes
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......