Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
77
If the field in the column "Pump" has a
black background, the entry mode for the
Pump level
is obtained by pressing the
key
.
With
or
select Pump Level 5
– 8
or
“-------” and confirm with
.
“------” stands for "no change to previous
segment", i.e. when
“------” is present in
all fields, the pump level always retains
the start setting or the setting before the
program start.
Continue with
to the field "Out 1",
"Out 2" or "Out 3".
The contact outputs of the contact
module (if present, special accessory)
are programmed here.
If the field in the column "Out 1" has a
black background, the entry mode for the
Contact output
is obtained by pressing
the
key.
With
or
select
-------
,
Open
or
Closed
and confirm with
.
-------
stands for no change with respect
to the previous segment, i.e. if
-------
is
present in all fields, the contact setting of
the start setting or that from the program
start is retained.
If applicable, continue with
to "Out 2"
and "Out 3".
Programming is terminated with
or
End .
Pump Menu
End
T
set
T
fix
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1
-------
Pump level
Pump Menu
End
T
set
T
fix
Contact out
-------
open
closed
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......