Proline Kryomats
80
Starting up
YAUE0007 / 10/2/2018
7.9
Ramp function
With the ramp function, temperature changes over any time period can be conveniently entered. This is
especially advantageous with very low temperature changes (e.g. 0.1 °C/ day).
Example: From the current outflow temperature (e.g. 242.4 °C) 200 °C of cooling is to occur over 5
days.
Then 200 °C is entered as the temperature change, the time value 5 is entered for the time and day(s)
selected as the time unit.
The ramp function is executed until it is manually terminated or until the temperature limits
T
il
(min) or T
ih
(max) described in Section 7.6.2 are attained.
Command
Ramp function
Open the list of device parameters using
the soft key
Menu .
With the cursor keys continue to
Programmer
Ramp function
.
Confirm with the key
.
Enter a positive or negative temperature
value with Temp. change .
With Time enter a figure (without time
unit).
With Time unit choose between
Second(s)
up to
Day(s)
.
Under Status the ramp is started
Start
or stopped
Stop
.
When the ramp function is being
executed,
Ramp active
appears in the
window bar.
Without manual switch-off the ramp
terminates at the latest at T
il
(min) or T
ih
(max).
Pump Menu
End
T
set
T
fix
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Program 1
Program 2
Program 3
Program 4
Program 5
Ramp function
Pump Menu
End
T
set
T
fix
Status
Temp. change
Time
Time unit
Second(s)
Minute(s)
Hour(s)
Day(s)
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......