Starting up
YAUE0007 / 10/2/2018
Command
Control Parameters
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to:
Control
Control Parameters
Control Parameters
.
The adjacent window appears.
Parameter endings:
e = Master controller, f = Slave controller.
Where applicable change parameters
marked with (auto) to manual input with
Tv manual/auto
.
Select the parameters to be changed with
and confirm with
.
Then change the value in the following
settings window and confirm with
.
Correction limitation
see introduction.
7.11.2.1
Steps for setting the control parameters for external control
1. Activate external control
2. Setting the slave controller:
2.1. Set parameters to
auto
; Check for thermostat type and change when necessary (RP....)
Choose heat transfer liquid with low viscosity and high thermal capacity.
Ranking: water, ethanol, water-glycol, oil, Fluorinert®.
Set pump level as high as possible,
make bath circulation strong and fast,
choose hose length as short as possible, i.e. 2 x 1m,
choose hose cross section as large as possible, i.e. ½ inch,
throughput through the external load as large as possible.
2.2. Xpf setting:
when oscillating with short period occur (i.e. 30 seconds)
Xpf lower, otherwise higher,
in case of bad thermal coupling and large thermal mass
high (i.e. 2
– 5, or even higher),
in case of good thermal coupling and small thermal mass
low (i.e. 0.2
– 0.7),
when rapid temperature response is required simple internal control should be preferred.
Otherwise select small Xpf (0.05
– 0.1).
3. Setting the master controller (PIDT1-controller):
Start with setting Auto and proceed with Manual only when necessary.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Kpe
0,50
Tne
100
Tve (auto) 83
Tde (auto)
8,3
Xpf
4,0
Prop_E(a) 30
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......