Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
89
7.12 Alarms, Warnings and Errors
The SelfCheck Assistant of your Proline Kryomat monitors more than 50 device parameters and triggers
alarms, warnings or errors as appropriate.
All warnings and alarms are shown on the Command remote control in plain text. Errors are shown in
plain text on the Command remote control, also, in an error list.
Alarms
: Alarms are safety relevant. Pump, heater and refrigerating unit will be shut off.
Warnings
: Warnings normally are not safety relevant. The device continues to operate.
Errors
:
If an error occurs, the pump, heater and refrigerating unit switch off automatically.
Switch of the unit at the rotary mains switch. If the error is always present after switching on
the device, please give information to the LAUDA Service for Constant Temperature
Equipment (
Find cause of alarm or warning and rectify where necessary. Then press
on the Master keyboard in
order to remove the alarm message. Warning messages can be removed either on the Master keyboard
with
or on the Command board with
.
Warnings may be ignored by pressing
or
on the Master keyboard or by activating the Screen
Softkey on the Command remote control. Warnings will not be repeated periodically.
7.12.1
Overtemperature protection and checking
W303
The units are designed for operation with non-flammable and flammable liquids to
DIN EN 61010-2-010.
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......