Proline Kryomats
94
Starting up
YAUE0007 / 10/2/2018
7.12.6
Pump-motor supervision: Dry running
The SelfCheck Assistant monitors the Varioflex pump:
1. Alarm sounds as dual-tone signal when the pump runs without
liquid. This can only occur when the float level measurement has
failed.
2. The display of
PuLEU
signals that the SelfCheck Assistant
has detected a pump low level.
3. The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Pump and refrigerating unit are switched off electronically.
The cause of the failure of the level measurement with the floatation sensor must be
found and rectified. Perhaps foreign bodies in the bath block it.
Press the Enter key.
Also press this key if the unit has been switched off in the fault
state.
Command
Alarm! Low level (pump)
Alarm! Low level (pump) is shown in the display and signifies
that unlocking is only possible on the Master control panel.
7.12.7
Compressor Overtemp
Command
Error! Comp1 overtemp
Error! Comp1 overtemp is shown in the display. The cause may
be a technical malfunction or an extremly situation in temperature
control. Switch off the unit and wait min. 15 minutes to restart it
again so that the compressor has time enough to cool down again.
This error may affect stage 1 (Error 68) or stage 2 (Error 69).
If the error is always present after switching on the device, please
give information to the LAUDA Service for Constant Temperature
Equipment (
9.5).
7.12.8
Three-phase current
Command
Error! Three-phase current
Error! Three-phase current is shown in the display, signed by
number 70. The cause is the wrong direction of the current rotation
field. It has to be clockwise!
Another reason may be the missing of one phase of the voltage
supply.
Pump alarm
PuLEU
°C
Summary of Contents for RP 3050 C
Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Page 2: ......
Page 4: ......
Page 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Page 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Page 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Page 134: ......
Page 135: ......